Wiring The Download Cables

The Diagrams below show you how to wire the download cable for use with your PAT Tester.

The Fluke and Seaward Supernova testers seem to use a 'Null Modem' serial cable.
The Metrel and Transmille Pat test machines all seem to use a straight through cable.
The Megger PAT4 has its own proprietary cable which is unique.


NOTE!
YOU MAY CAUSE DAMAGE TO YOUR PAT TESTER OR COMPUTER IF YOU USE THE WRONG DOWNLOAD CABLE!

The tables show the wiring of the PINs in a 9pin to 9pin cable. Male end connecting to the PAT and Female end connecting to the PC.

Metrel AlphaPat, BetaPat, OmegaPat and SigmaPat, Transmille 6080 and 6080 PLUS

 

Signal

Omega

PC

 

DCD

1

1

 

RXD

2

2

 

TXD

3

3

 

DTR

4

4

 

GND

5

5

 

DSR

6

6

 

RTS

7

7

 

CTS

8

8

 

+5V

9

9

Seaward Supernova (including PLUS & ELITE), Europa, Europa+, ETHOS 9500, Fluke 6500 Pat Testers

  Signal Europa PC
  DCD    
  RXD 2 3
  TXD 3 2
  DTR   4, 6 and 8 Joined
  GND 5 5
  DSR   4, 6 and 8 Joined
  RTS    
  CTS   4, 6 and 8 Joined
  +5V    

Megger PAT4 DVe

  Signal PAT4 PC
  DCD 1 7
  RXD 2 3
  TXD 3 2
  DTR 4 6 and 8
  GND 5 5
  DSR 6 4
  RTS 7 1
  CTS 8 4
  +5V 9 9

Robin SmartPat 5000, Seaward 1000X, 1000s, IT1000, PAT1000

  Signal SmartPat PC
  DCD    
  RXD 2 4 and 6 and 8
  TXD 3 2
  DTR    
  GND 5 5
  DSR    
  RTS    
  CTS    
  +5V    

Kewtech KT74, KT75 and KT77

  Signal SmartPat PC
  DCD    
  RXD 2 4 and 6 and 8
  TXD 3 2
  DTR    
  GND 5 5
  DSR    
  RTS    
  CTS    
  +5V    

Martindale MPATPLUS (MicroPat+), Metrotest MPAT40 and MPAT60

Martindale Part Number TL125

  Signal MicroPat+ PC
  DCD    
  RXD 2 4 and 6 and 8
  TXD 3 2
  DTR    
  GND 5 5
  DSR    
  RTS    
  CTS    
  +5V